"Sample high pressure limit" error when priming wash syringes on ACQUITY - WKB11769
Article number: 11769
SYMPTOMS
- Get a sample overpressure error message when priming wash syringes
- Can prime without error after disconnecting the Volume Detection Device (VDD; air sensor, bubble detect device)
- Characterization of the bubble detect device fails
ENVIRONMENT
- ACQUITY
CAUSE
Faulty Volume Detection Device (VDD)
FIX or WORKAROUND
- Replace the Volume Detection Device (700002867).