ACQUITY Arc Sample Manager FTN-R: What is the required extension loop and syringe size for various injections? - WKB116885
Article number: 116885
ENVIRONMENT
- ACQUITY Arc Sample Manager FTN-R
ANSWER
For 100 µL injection, a 250 µL expansion loop should be installed.
It is possible to inject 100μL with a 100μL syringe and 100uL sample loop.
However, it is a good practice to have the syringe size at least 2x the volume of sample loop.



ADDITIONAL INFORMATION
Extension Loop Options:
430002585: Kit, Extension Loop, MP35N, 1000uL

Syringe Options:
- Syringe, 50 µL (# 410001348)
- Syringe, 100 µL, HP (# 700002570)
- Syringe, 250 µL (# 410001347)
- Syringe, 500 µL, HPLC (# 410001647)
Be sure to configure the correct volumes in ACQUITY Console, How to confirm the ACQUITY UPLC H-Class loop size/needle size/sample syringe size from the Console screen? - WKB87507