Skip to main content
Waters

ACQUITY Arc Sample Manager FTN-R: What is the required extension loop and syringe size for various injections? - WKB116885

Article number: 116885

ENVIRONMENT

  • ACQUITY Arc Sample Manager FTN-R

ANSWER

For 100 µL injection, a 250 µL expansion loop should be installed.
It is possible to inject 100μL with a 100μL syringe and 100uL sample loop.

However, it is a good practice to have the syringe size at least 2x the volume of sample loop.

clipboard_e95bbdaeffaa358108404bb2e5b2ac929.png

clipboard_e0915a69dbdd1deab0e9da19ce0d4f607.png

clipboard_e507b41cc7aba9c5a8ad486f28a0d0b19.png

 

ADDITIONAL INFORMATION

Extension Loop Options:

430002585: Kit, Extension Loop, MP35N, 1000uL 

clipboard_effd8b0a19ecdd7bdc6748615eb15a39f.png

Syringe Options:

  • Syringe, 50 µL (# 410001348)
  • Syringe, 100 µL, HP (# 700002570)
  • Syringe, 250 µL (# 410001347)
  • Syringe, 500 µL, HPLC (# 410001647)

Be sure to configure the correct volumes in ACQUITY Console, How to confirm the ACQUITY UPLC H-Class loop size/needle size/sample syringe size from the Console screen? - WKB87507

id116885, UPAQSM, UPARBQSM, UPARBSMFTN, UPARCH30, UPARHQSM, UPARQSM, UPARSMFTN, UPPDAARCB