ACQUITY H-Class QSM fails gradient test - WKB116904

Article number: 116904

SYMPTOMS

  • ACQUITY H-Class QSM gives bad gradient profile
  • clipboard_e1b23e23260038da9e2a07005b2f3c877.png

ENVIRONMENT

  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

CAUSE

The solvent in-line filter is blocked on the D line.

FIX or WORKAROUND

Replace or clean the solvent in-line filter (700003615 - Filter, Solvent Bottle, SS PKG 1 or 700003616 - Filter, Solvent Bottle, SS PKG 7).

ADDITIONAL INFORMATION