Contamination peak on system - WKB114534
Article number: 114534
SYMPTOMS
- Contamination peak present in THC trace for solvent blanks.
- Contamination increases after subsequent injections.
- Customer replaced all components of the injection pathway (needle, seal, seal housing) with no improvement.
ENVIRONMENT
- Xevo TQ-S
- Xevo TQ-XS
- ACQUITY I-Class
CAUSE
Residue left over from the vial septa was reacting with something from the nitrogen of a peak generator.
FIX or WORKAROUND
Nitrogen from a different generator removed the contamination peak. Peak engineer needed to service or replace the generator.
ADDITIONAL INFORMATION
Waters TruView vials also resulted in far less of a contamination peak than the standard blue lidded vials, so must produce less of the septa residue that reacts with the nitrogen.