No peaks eluting from the GC column when using APGC - WKB114229
Article number: 114229
SYMPTOMS
- No peaks elute from column
- Complete noise background; no bleed peaks come off
- Cone gas set to 0 in order to elevate any background peaks (should be between 100 and 150 in stepwave instrument)
- Peaks ionizing the heptacosa in background
- All other source conditions tuned normally
ENVIRONMENT
- APGC
- UNIFI
- Xevo G2-XS
- Xevo G2-S
- Xevo TQ-S
- Xevo TQ-XS
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
CAUSE
Broken column (column snapped in the middle)
FIX or WORKAROUND
Swap columns
ADDITIONAL INFORMATION