Alliance 2695 Sample Heater/Chiller will not cool to set point - WKB11392
Article number: 11392
SYMPTOMS
- Sample heater chiller does not cool to the set point.
- Sample compartment is defrosted (by either setting temperature to 40 degrees for 2 hours or powering off Alliance overnight).
- Room temperature is normal with no fluctuations
- No heat source near instrument
- Fan at rear of instrument is air flowing from both sides.
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Faulty heater cooler engine.
FIX or WORKAROUND
Replace the heater/cooler engine - part number 700001066
ADDITIONAL INFORMATION
Very old 2690 systems might also need a heater/cooler foam kit 700001295
Part number for Alliance Sample heater/cooler assay: WAT270804
Check and confirm that the power supply is giving a correct voltage output. See Document 715004034.