Skip to main content
Waters

2695 "lost prime" error - WKB113750

Article number: 113750

SYMPTOMS

  • "Lost prime" error message on panel

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • 2795 Separations Module

CAUSE

Large salt buildup on fittings from the vent valve to the inline filter

FIX or WORKAROUND

Replace the tubing. (Assembly, Tubing Vent Left-In Line #WAT270974).

ADDITIONAL INFORMATION

Have the filter (WAT035190) on hand as well.

 

 

Not able to find a solution? Click here to request help.