Article number: 113161

SYMPTOMS

  • All tubings are submerged in solvent
  • Tubings are filled with solvent
  • Disconnecting B1/B2 from the solvent check valve brings dripping
  • When priming pump B, no solvent comes out

ENVIRONMENT

  • ACQUITY UPLC BSM

CAUSE

The i2V check valve is clogged.

FIX or WORKAROUND

  1. Replace the i2V check valve.
  2. Prime the system.
  3. Perform the dynamic leak test.

ADDITIONAL INFORMATION