Instrument method errors with flow too high (>1.23) for stroke volume when setting up an Alliance-based LC system - WKB10956
Article number: 10956
SYMPTOMS
- Instrument failure in the Empower Run Samples window
ENVIRONMENT
- Empower
- Empower 2
- Empower 3
CAUSE
The flow rate in the instrument method is outside the range for the current stroke volume setting.
FIX or WORKAROUND
- Set the "Default" method current via the front panel on the Alliance. (This will reset the stroke volume range.)
- Re-engage the target instrument method via Empower.
ADDITIONAL INFORMATION
id10956, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP