What routine maintenance is recommended for my 2465 electrochemical detector? - WKB109555
ENVIRONMENT
- 2465 Electrochemical Detector
- 2695 Separations Module
ANSWER
Consult Chapter 4 of the 2465 Operator's Guide (71500246502)
ADDITIONAL INFORMATION
The operator's guide includes the following tasks, which can help improve chromatography and sensitivity issues:
1. Perform HPLC system passivation.
2. Clean the working electrode.
3. Maintain the reference electrode.
4. Clean the HPLC column per the manufacturer's care and use guide.
5. Use only HPLC-grade water.
6. Use the correct flow cell spacer. If any maintenance requires you to disassemble the flow cell, the spacer must be replaced with a new one. It shouldn’t be reused.
7. If the HPLC system is idle for more than 48 hours, remove any buffers from the system and flush with water. Use 50:50 water/methanol for storage.
Returning the system back to buffers calls for the reverse—prime with water and then with buffers.
8. If you are not using your HPLC column for a similar period, it should be flushed following the manufacturer's care and use guide—typically, flush off buffers with water and then store in water/methanol. If you are changing the column, note the requirement to flush at low flow for 3 days.
9. A periodic flow cell flush is beneficial—30 minutes at 0.5 mL/min with water or ethanol with column removed.
10. To prevent formation of air bubbles in the flow cell, the flow cell should be positioned at a 45-degree angle, and the waste line outlet tubing from the flow cell should run to a waste container that is at a higher level than the flow cell. This gives a minute back pressure that prevents air-bubble entrapment.
The outlet tubing should be under the liquid level to prevent electrical baseline noise.
id109555, 2465, baseline ripple