Loss of sensitivity on 2465 Electrochemical Detector - WKB109553

Article number: 109553

SYMPTOMS

  • Loss of sensitivity on 2465 Electrochemical Detector
  • Replacing the HPLC column made no significant difference
  • Passivating the HPLC system made no significant difference
  • Cleaning the working electrode made no significant difference

ENVIRONMENT

  • Empower 
  • 2465 Electrochemical Detector
  • 2695 Separations Module

CAUSE

The wrong flow cell spacer was installed.

ADDITIONAL INFORMATION