Loss of sensitivity on 2465 Electrochemical Detector - WKB109553
Article number: 109553
SYMPTOMS
- Loss of sensitivity on 2465 Electrochemical Detector
- Replacing the HPLC column made no significant difference
- Passivating the HPLC system made no significant difference
- Cleaning the working electrode made no significant difference
ENVIRONMENT
- Empower
- 2465 Electrochemical Detector
- 2695 Separations Module
CAUSE
The wrong flow cell spacer was installed.
FIX or WORKAROUND
Install the correct flow cell spacer. See What are the different options for spacers for a 2465 Electrochemical Detector flow cell? - WKB23564
ADDITIONAL INFORMATION