Status message on front panel of Alliance 2695: A hardware fault has been detected by the column heater electronics - WKB107701

Article number: 107701

SYMPTOMS

  • Message on the front panel of Alliance 2695: "A hardware fault has been detected by the column heater electronics"
  • Reboot does not resolve
  • Cannot set column heater temperature

column hw fault.JPG

 

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2695D Dissolution Separations Module
  • Alliance Series Column Heater
  • Alliance Series Column Heater/Cooler

CAUSE

The cable is not seated properly.

FIX or WORKAROUND

  1. Power-off the 2695 and the column heater.
  2. Reseat the cable between the 2695 and the column heater.
  3. Power-on the 2695 and the column heater.

ADDITIONAL INFORMATION

 