2487 Dual-Wavelength Absorbance Detector lost communication - WKB107796
Article number: 107796
SYMPTOMS
- 2487 Dual-Wavelength Absorbance Detector lost communication
- "Instrument Failure" in message center
- Status NO in Node properties
- Serial number missing from instrument display
ENVIRONMENT
- 2487 Dual-Wavelength Absorbance Detector
- Empower 3
CAUSE
The battery was replaced, causing the IEEE address to default to 4 and the serial number to be missing. This caused a conflict with another 2487 detector with the same IEEE address of 4.
FIX or WORKAROUND
Re-enter the serial number from the front panel and set the IEEE address to 5, or any address that is not used by other IEEE equipment.