ACQUITY ELSD gives a noisy baseline and then a flat baseline - WKB107634
Article number: 107634
SYMPTOMS
- At a gain of 1, when flow applied, baseline is erratic, between 0 and 900 LSU
- At a gain of 100, flat baseline
- With no flow for 30 minutes to clear drift tube, normalization is 2.1
ENVIRONMENT
- ACQUITY UPLC ELS Detector
CAUSE
Extract exhaust is not working sufficiently.
FIX or WORKAROUND
Ensure that the extract exhaust is set to the correct flow.
ADDITIONAL INFORMATION
There should be a minimum negative pressure of −0.21 kPa (−0.00 bar, −0.03 psi) between the detector and the laboratory exhaust system.
id107634, UPELS, UPELSARC