Skip to main content
Waters

Peptides not resolved (ACQUITY M-Class) - WKB10708

Article number: 10708

SYMPTOMS

  • Peptides come off the column not resolved
  • Hydrophobic peptides elute together
  • Hydrophyllic peptides elute later than usual and not resolved

ENVIRONMENT

  • ACQUITY UPLC M-Class

CAUSE

  • Flow controller needs to be re-zeroed
  • Setting for "Loop Offline" parameter is incorrect

FIX or WORKAROUND

  1. Perform "Auto Zero Flow Transducers" procedure for both pumps A and B.
  2. Verify that the "Loop Offline" setting for M-Class Sample Manager parameters in the Inlet Method is either set to "Disabled" or there is sufficient time to allow complete sample load out of the loop and onto the column.

ADDITIONAL INFORMATION

Prior to performing the "Auto Zero Flow Transducers" procedure, ensure that the pump has been flowing 50% A : 50% B for some time to prevent backflow of solvents and incorrect results of the procedure.

https://support.waters.com/KB_Inst/C...ow_transducers

Not able to find a solution? Click here to request help.