"Needle seal force sensor h/w fault (0.0)" error on ACQUITY UPLC Sample Manager - WKB107057
Article number: 107057
SYMPTOMS
- Error messages:
- needle/seal move limit exceeded
- Z-axis calibration out of range
- Seal force sensor h/w fault (0.0)
ENVIRONMENT
- ACQUITY APC Sample Manager - FTN
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-Class
CAUSE
Bent needle.
FIX or WORKAROUND
Replace needle.
ADDITIONAL INFORMATION
For information on how to replace the needle:
id107057,
