Ferrule stuck inside injector valve port - WKB10654
Article number: 10654
SYMPTOMS
- Ferrule is stuck inside one of the injector valve ports (typically port 5 or 6)
ENVIRONMENT
- ACQUITY APC
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC M-Class
- nanoACQUITY UPLC
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
- ACQUITY UPLC CM-A
CAUSE
Over-tightened tubing.
FIX or WORKAROUND
- Attempt to remove ferrule without damaging injector valve pod
- If unsuccessful, replace injector valve pod
- Verify integrity of fluidics connections by performing Set Pressure Diagnostics test (M-Class and nanoACQUITY) or Dynamic Leak Test including seal and needle
ADDITIONAL INFORMATION
Also similar issue can be resolved on the Acquity column manager but with a 7 port valve