Poor injection reproducibility with Alliance 2695 - WKB10442

Article number: 10442

SYMPTOMS

  • System is not delivering reproducible injections
  • Inconsistent area counts between injections from the same vial
  • Poor system precision
  • Poor peak shape, tailing, etc.

ENVIRONMENT

  • Alliance HPLC system
  • 2695 Separations Module

CAUSE

Air is leaving the system through the purge waste line, attributable to the seal pack being out of adjustment/alignment.

FIX or WORKAROUND

  1. Perform a wet prime.
  2. Verify that no air bubbles are leaving the system through the purge waste line.
  3. Perform "Adjust Seals" from front panel of the 2690/2695.
  4. Note any change from "Previous" to "Current" values.
  5. Verify that the values are within the recommended Waters range (see Additional Information section below).
  6. Accept the values.

ADDITIONAL INFORMATION