Peak saturation on 2475 detector in low-sensitivity mode - WKB10406
Article number: 10406
SYMPTOMS
- Peak saturation after repair service
- Peaks off scale
- Flat-top peaks
ENVIRONMENT
- 2475 detector
- It is impossible to change the method (reduce injection volume and gain setting, for example)
CAUSE
Sensitivity is recovering with repair service (repair optics).
FIX or WORKAROUND
To reduce PMT sensitivity:
-
Select Diag > 8 Other diagnostics on 2475.
-
Press "PMT Sensitivity override" and show the "PMT Sensitivity" view.
-
Select "Low Sensitivity Mode".
-
Specify a reduction value of 10 or 100.
ADDITIONAL INFORMATION
2475 Multi-Wavelength Fluorescence Detector Overview and Maintenance Guide
https://help.waters.com/help/en/support/library-details.html?documentid=715004756