SYMPTOMS

  • Dynamic leak test fails

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class

CAUSE

Dirty check valve

FIX or WORKAROUND

Sonicate or replace the check valve.