ACQUITY UPLC I-Class fails dynamic leak test - WKB101654
Article number: 101654
SYMPTOMS
- Dynamic leak test fails
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC I-Class
CAUSE
Dirty check valve
FIX or WORKAROUND
Sonicate or replace the check valve.
ADDITIONAL INFORMATION
id101654, checkvalve, DLT, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO