No signal from the 2489 detector in Empower 3 - WKB10162
Article number: 10162
SYMPTOMS
- 2489 shows "yes" for status ok in the node properties
- 2489 instrument control panel display works, lamp icon is green
- 2489 stays in a ready state when injection occurs
ENVIRONMENT
- 2489 UV/Visible Detector
- 1525 Binary HPLC pump with manual injector
- Empower 3 stand-alone
CAUSE
I/O connections in the wrong port.
FIX or WORKAROUND
- Inject start connections go in the II connection in the rear of the 2489.
ADDITIONAL INFORMATION
id10162, 1525, 1525U, 1525UIVD, 2489, 2489B, 2489SFC, A-10UV, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP