Skip to main content
Waters

No signal from the 2489 detector in Empower 3 - WKB10162

Article number: 10162

SYMPTOMS

  • 2489 shows "yes" for status ok in the node properties
  • 2489 instrument control panel display works, lamp icon is green
  • 2489 stays in a ready state when injection occurs 

ENVIRONMENT

  • 2489 UV/Visible Detector
  • 1525 Binary HPLC pump with manual injector
  • Empower 3 stand-alone

CAUSE

I/O connections in the wrong port.

FIX or WORKAROUND

  1. Inject start connections go in the II connection in the rear of the 2489.

  2489 Inject start.JPG

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.