2475 detector does not start acquisition after injection - WKB100212
Article number: 100212
SYMPTOMS
- Acquisition does not start after injection
- Battery has been replaced
- Status is OK yes
- Controlled by RS232 cable
ENVIRONMENT
- 2475 detector
CAUSE
The 474 emulation mode configuration is not selected.
FIX or WORKAROUND
- Configure the detector for 474 emulation.
- Select CONFIGURE (Shift, DIAG).
- The first of three Configuration screens appears.
- Select "474 emulation".