Skip to main content
Waters

Why do waters_connect apps report inconsistent dates and times for acquisitions? - WKB257093

Article number: 257093

SYMPTOMS

  • Waters_connect does not report times in the same format across all apps
  • The times recorded are correct, but not in the same time zone; some will be local time, some will be UTC

ENVIRONMENT

  • waters_connect for Quantitation 1.0
  • waters_connect for Quantitation 1.1
  • MS Quan 1.4.0
  • MS Quan 1.7.0
  • UNIFI 2.1.1
  • UNIFI 3.1.0

CAUSE

Acquisition times recorded into the database use UTC (+00:00) instead of the local time. Newer applications, such as MS Quan and LC-MS Toolkit, might display the Acquisition dates and times with a time offset for records such as result sets, but not for injection items.

FIX or WORKAROUND

None

ADDITIONAL INFORMATION

CRI-5414 recorded. This behavior does not impact data integrity.

id257093,

Not able to find a solution? Click here to request help.