WDM backups fail with error ORA-27369 and exit code "Incorrect function" - WKB95268
Article number: 95268
SYMPTOMS
- WDM backups fail despite reinstallation of WDM and permission checks for oraclejobs
- ORA-12012: error on auto execute of job "WATERS"."FIRST_FULL_LVL0"
- ORA-27369: job of type EXECUTABLE failed with exit code: Incorrect function
- Backup reports did not have date appended to the backup job
ENVIRONMENT
- UNIFI 1.9.4 network
CAUSE
Powershell was blocked by AV (Cyclance) when the script tried to run the command to append the time/date.
FIX or WORKAROUND
Request IT to add an exception to the AV so that it ignores powershell scripts running from the UNIFI directories.
ADDITIONAL INFORMATION