Unable to bring LND online - WKB94931

Article number: 94931

SYMPTOMS

  • After the PSU was replaced on the LND, no communications apparent; "Unable to bring LND Online"
  • EPC cannot be pinged from the LND

ENVIRONMENT

  • VION IMS QTof
  • UNIFI Network

CAUSE

The Ethernet cable from the EPC is plugged into the wrong port on the LND (there are two unbridged ports)

FIX or WORKAROUND

Connect the Ethernet cable from the EPC to the correct port.