UNIFI does not make injection - WKB94232
Article number: 94232
SYMPTOMS
- UNIFI system starts acquisition but never makes injection
- If pre-injection action is specified in the method, status sits on Pre-Injection Action and remaining time never advances
- If pre-injection action is not specified, status says Injection Started and remaining time counts down, but no injection takes place and no data is acquired
ENVIRONMENT
- UNIFI 1.9.4 workstation
- Xevo G2-XS QToF
- I-Class BSM, FTN, PDA
CAUSE
Unknown
FIX or WORKAROUND
- Stop and close analysis.
- Close UNIFI portal.
- Reboot workstation PC.
- Wait for Waters.ServiceHost.exe instances to start.
- Reboot EPC using wt_Reboot command over serial connection or telnet.
- Start portal and log in.
- In Administration > Device Management, right-click on the workstation and choose Reset. Say Yes at the prompt.
- Wait for the workstation status to say Online.
- Make sure the Instrument System is Online.
- Open the system in My Work.
- Put MS in Operate and turn on the lamp if necessary.
- Create new analysis.
- Bring up System side panel in the Analysis window and choose Go To Initial conditions from System pane drop-down menu.
- Wait for system to stabilize.
- Submit samples.