Where is the DHCP server configuration file in UNIFI? - WKB840

Article number: 840

ENVIRONMENT

  • UNIFI
  • waters_connect
  • LND

ANSWER

Find the DHCP Server wizard at install drive of LND

UNIFI

C:\Program Files\Waters\UNIFI\Server\Bin\

waters_connect

C:\Program Files (x86)\Waters\LND\DHCP\

ADDITIONAL INFORMATION

The DHCP file is located here
[UNIFI install drive]:\Program Files\Waters\UNIFI\Server\Bin\DHCP.XML

w_c

C:\Program Files (x86)\Waters\LND\DHCP\DHCP.xml

NB. Never delete or edit the DHCP.xml in UNIFI. The information is mirrored in the database and deleting is likely to cause conflicting entries and a stuck offline LND or Workstation.