Calibration curve is not produced in UNIFI Quantify Assay TOF 2D Chromatographic method - WKB78939

Article number: 78939

SYMPTOMS

  • Standard compound is identified correctly in all injections, but the calibration curve is not generated

ENVIRONMENT

  • UNIFI 1.9.4.053

CAUSE

The levels selected in "sample information" do not match the ones in the "default amount" column of the Purpose tab of the method.

   clipboard_ead7648efe88c19b1db4d20d1ffcb5c13.pngclipboard_eadf88e494c716502e29810aa9049705d.png

FIX or WORKAROUND

Change the levels in "sample information editor" to Level 1, Level 2, ....

clipboard_e62f7ab732eec1bce85a0fb6b99a13229.png

ADDITIONAL INFORMATION

 

id78939, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC