Explorer is blank / not populated (CRI-465) - WKB74308
Article number: 74308
SYMPTOMS
- A user is unable to view folders; when closing Explorer the following error appears:
"Cannot update user preferences' window:
An error occurred while saving the home node, if the problem persists contact your System administrator"
ENVIRONMENT
- UNIFI 1.9.4
CAUSE
CRI-465
When using Explorer to frequently search across many folders, the user preferences can become corrupted and result in this defect.
FIX or WORKAROUND
- Contact Waters Support for excalation to Waters Global Service Support for versions 1.9.9 and earlier
- The defect is now resolved in waters_connect UNIFI 1.9.12 and above due to a fix in an Oracle Defect in 19c
ADDITIONAL INFORMATION
id74308,