UNIFI calibration fails after 1.9.4 upgrade - WKB72758

Article number: 72758

SYMPTOMS

The calibration appears to start, but stays on "Calibration Setup Pending" for a few minutes, and then goes back to idle.

ENVIRONMENT

  • UNIFI 1.9.4
  • Xevo G2XS QTof

CAUSE

Using calibration profiles created before the upgrade causes issues.

FIX or WORKAROUND

Create a new calibration profile with the same parameters as the old one.

ADDITIONAL INFORMATION