"Probe not connected" error for Xevo instrument - WKB72679

Article number: 72679

SYMPTOMS

  • In UNIFI/MassLynx, "probe not connected error" appears, sometimes with "Idle 71570338.57 Minutes Error" as additional information

ENVIRONMENT

  • UNIFI 1.9.4
  • MassLynx 4.2
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS QTof

CAUSE

Loss of communication

FIX or WORKAROUND

  1. Reboot the electronics on the mass spectrometer.
  2. Restart the computer.

ADDITIONAL INFORMATION

Here is the procedure for rebooting the electronics on a Xevo G2.