Cannot save an analysis in UNIFI - WKB64619

Article number: 64619

SYMPTOMS

  • When trying to save an analysis, an error appears saying that the process data is not in a consistent state
    • Software prompts user to reprocess, but unable to find the process button    

ENVIRONMENT

  • UNIFI 1.9.4 (workstation)

CAUSE

Looking on the wrong tab within the Analysis 

FIX or WORKAROUND

  1. Switch tabs within the UNIFI analysis from Investigate to Review.
  2. The Process menu appears and reprocessing can continue.

ADDITIONAL INFORMATION