Microsoft Access error when trying to install Office on a UNIFI PC - WKB62377

Article number: 62377

SYMPTOMS

  • Tried to install Microsoft Office software, but could not install it because Microsoft Access is already installed
  • UNIFI 1.9.4 (or other version of UNIFI) is installed, but neither the engineer nor the customer have installed Microsoft Access
  • MS Database Access Engine 2010 is present in Windows Control Panel > Programs and Features

ENVIRONMENT

  • UNIFI 1.9.4
  • Likely affects other versions of UNIFI as well (see Cause)
  • MS Office
  • Windows 7
  • Windows 10

CAUSE

UNIFI data export to Excel requires that the 32-bit version of the MS Database Access Engine 2010 be installed on the PC. This is installed from the UNIFI DVD as a prerequisite.

FIX or WORKAROUND

  1. Remove the 32-bit version of the MS Database Access Engine 2010 from the PC.
  2. Install Microsoft Office.
  3. Check for 64-bit version of MS Database Access Engine 2010, and remove it if necessary (see Additional Information).
  4. Reinstall the 32-bit version of the MS Database Access Engine 2010 from the prerequisites folder of the UNIFI installation DVD.

ADDITIONAL INFORMATION

If the 32-bit version of the MS Database Access Engine 2010 is removed by an installer, overwritten by the 64-bit version of MS Database Access Engine 2010, or a newer version of MS Database Access Engine is installed, that will break UNIFI's "Export to Excel" functionality. In that case, follow steps 3 and 4 above.

Not able to find a solution? Click here to request help.