How do you apply Microsoft security updates in UNIFI - WKB62224

Article number: 62224

ENVIRONMENT

  • UNIFI 1.9.4 Hot Fix 1 only
  • Server or Workstation

ANSWER

  1. Determine the Waters testing results for UNIFI 1.9.4 Hot Fix 1 servers. (What are the UNIFI 1.9.4 Microsoft security update testing results?)
  2. Ensure that all users are logged out.
  3. Ensure that there are no running analyses or processing jobs.
  4. Apply the Microsoft Security updates. Do not apply optional updates, especially those for .NET (.NET security updates can be applied).
  5. Reboot the server after updates are completed.
  6. Confirm that UNIFI functionality is unaffected.
  7. If problems are observed, remove the updates and contact Waters Support for advice.

ADDITIONAL INFORMATION