Intermittent instrument communication issues - WKB54543

Article number: 54543

SYMPTOMS

  • Instrument disconnects when idle for some time and application window is unresponsive

ENVIRONMENT

  • UNIFI v1.9.3 Network
  • LND running Windows 7
  • MassLynx
  • Empower

CAUSE

The power setting on the PCI Express slot is set to "Maximum power savings".

clipboard_eaee1812e25cb113d22949eda98784989.png

FIX or WORKAROUND

On the Advanced settings tab, change the PCI Express setting to Off.

clipboard_e4b9ea472d13e66e150fcf503aa6d225a.png

ADDITIONAL INFORMATION

This PCI Express power setting manages power settings for the PCI slot where the Ethernet Network Interface Card (NIC) is. This would also impact any NICs and BusLACE cards for LAC/Es on Empower and NICs and NI GPIB cards for MassLynx.