In UNIFI, get an "Error writing stream" when trying to import MassLynx data - WKB231010

Article number: 231010

SYMPTOMS

  • When trying to import MassLynx raw data from a local drive into UNIFI, the import doesn't complete and errors out
  • The Event Browser window shows an Acquisition File Data Import Error; the details show Error Writing stream: ORA-01691: unable to extend lob segment EVERST.CON_ITEMSTREAMBLOB_STREAMC_LOB by 1024 in tablespace WATERSSTREAMBLOB.  Example:ORA error1.png
  • The Event Browser window shows an Acquisition File Data Import Error; the details show Error: <insert file name> Error writing stream: Stream not open.  Example:error writing stream1.png
  • There's plenty of space in the database
  • Stopping and starting the WatServiceHost service does not resolve

ENVIRONMENT

  • UNIFI 1.9.4 workstation
  • MassLynx

CAUSE

Suspect power outage affected PC

FIX or WORKAROUND

  1. Close all programs.
  2. Reboot the UNIFI workstation.

ADDITIONAL INFORMATION

Can also occur when data acquisition is running . Reboot resolved

id231010, MLYNX, MLYNXV41, SUPMM, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC