Golden Rules for UNIFI PC Management V – Checking Data Backups - WKB208961
OBJECTIVE or GOAL
Follow best practices in managing data backups on a UNIFI or waters_connect UNIFI PC (workstation and LND).
ENVIRONMENT
- UNIFI workstation
- UNIFI network
- waters_connect UNIFI workstation
- waters_connect UNIFI network
- Windows 7 UNIFI PC
- Windows 10 UNIFI PC
PROCEDURE
To protect UNIFI data, you must back up the Oracle database on a regular basis. Backups protect against hardware corruption or failure, such as the failure of a database drive on a workstation, or database corruption from a virus or malware.
-
It is important to check on a weekly basis that a full back up of the database has been a success. This is done by viewing “Manage Backups” in the Waters Database Manager (WDM):
o Start WDM and log in as administrator.
o View “Manage Backups” > “Scheduled Backups”.
o Click on the Log file icon
for the FULL_LVL0 backup.o Determine when the last full backup was successfully completed.
-
If there is no evidence that the backup has been running successfully or if it has failed, immediately contact your Local Waters Service Support representative.
-
Ensure that the backup area is NOT on the same C:\ Drive or the UNIFI installed drive.
Link to the Golden Rules of UNIFI PC Management - Main menu.