Error ORA-22059 in UNIFI: buffer size [] is too small – [] is needed - WKB12877
Article number: 12877
SYMPTOMS
- While making a new Analysis method in the Expected Components section, searching for library entries and trying to add them to the Manage Components list generates this error: "ORA-22059: buffer size [] is too small – [] is needed".

ENVIRONMENT
UNIFI 1.8.2 (workstation configuration).
CAUSE
Corrupt library/library items.
FIX or WORKAROUND
- Purge corrupted library entries from the library (refer to How to Find Corrupted Library Entries in UNIFI for details), or delete the library (see Additional Information). Avoid this problem by ensuring that any libraries in use are exported prior to any UNIFI version upgrades.
- After the upgrade, import the libraries back into UNIFI.
ADDITIONAL INFORMATION
To delete a library:
- Go to the Administration page in UNIFI.
- Under Scientific Libraries, select Manage Libraries.
- Select the library to delete.
- Select Delete at the top of the window.