With UNIFI, settings specified in analysis method or sample list are not applied during the analysis - WKB118090
Article number: 118090
SYMPTOMS
- Settings specified in the analysis method or promotable parameters in the sample list are not applied during analysis
- The "acquisition information" screen shows that values from manual tune were applied during acquisition
ENVIRONMENT
- Xevo G2-XS QTof
- UNIFI 1.9.4
CAUSE
"Locked parameters" were mistakenly enabled in the instrument setup screen, which prevents selected parameters from being overridden by method settings or promotable parameters. This option should not be used under normal circumstances.
FIX or WORKAROUND
Clear the "locked parameters" selection, as shown in the screenshot above.