SYMPTOMS
- UNIFI is unavailable after applying Windows updates to server and clients
ENVIRONMENT
- UNIFI 1.9.4 Network
- UNIFI 1.9.4 Workstation
CAUSE
.NET 4.8 has been applied to workstation or server but .NET 4.7.2 is the supported version for UNIFI.
FIX or WORKAROUND
Remove the following updates:
-
KB4486153, which is the .NET 4.8 update for Win 10 build 1809, 1803, and 1709
-
KB4486129, which is the .NET 4.8 update for Win 10 build 1703 and 1607
-
KB4486105, which is the .NET 4.8 update for Server 2012r2
-
KB4601051 Cumulative Update for .NET Framework 4.8 for Windows 10, version 1607