ACQUITY Console plot is stuck - WKB67890
Article number: 67890
SYMPTOMS
- ACQUITY Console plot is stuck; no movement or real-time display
ENVIRONMENT
- ACQUITY Console
CAUSE
The real-time button in the ACQUITY Console was pressed, which turned off the real-time plotting.
FIX or WORKAROUND
Press the real-time button on the right-hand side of the ACQUITY Console.
ADDITIONAL INFORMATION