Skip to main content
Waters

Timebase error using Empower and Corona Veo charged aerosol detector (CAD) - WKB4124

Article number: 4124

SYMPTOMS

  • Message appears in Empower 3 Feature Release 2 when running the CAD: "Found the following errors while linking panel to timebase [system name]"
  • Error "property search failed because "connected" is not unique"

ENVIRONMENT

  • Empower 3 Feature Release 2
  • Thermo Fisher Scientific Dionex Corona Veo Charged Aerosol Detector (CAD)

CAUSE

Unknown

FIX or WORKAROUND

  1. Reboot the system and software.
  2. Create a new instrument method.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.