Timebase error using Empower and Corona Veo charged aerosol detector (CAD) - WKB4124
Article number: 4124
SYMPTOMS
- Message appears in Empower 3 Feature Release 2 when running the CAD: "Found the following errors while linking panel to timebase [system name]".
- Error "property search failed because "connected" is not unique".
- Injection not triggered
- Data may display when monitoring baseline
ENVIRONMENT
- Empower 3 Feature Release 2.
- Thermo Fisher Scientific Dionex Corona Veo Charged Aerosol Detector (CAD).
CAUSE
Unknown.
FIX or WORKAROUND
- Reboot the system and software.
- Create a new instrument method.
- Check the RS232 to screw terminal connector cable
ADDITIONAL INFORMATION
