Timebase error using Empower and Corona Veo charged aerosol detector (CAD) - WKB4124

Article number: 4124

SYMPTOMS

  • Message appears in Empower 3 Feature Release 2 when running the CAD: "Found the following errors while linking panel to timebase [system name]".
  • Error "property search failed because "connected" is not unique".
  • Injection not triggered
  • Data may display when monitoring baseline

ENVIRONMENT

  • Empower 3 Feature Release 2.
  • Thermo Fisher Scientific Dionex Corona Veo Charged Aerosol Detector (CAD).

CAUSE

Unknown.

FIX or WORKAROUND

  1. Reboot the system and software.
  2. Create a new instrument method.
  3. Check the RS232 to screw terminal connector cable

ADDITIONAL INFORMATION

 