AAO Service error in Analyst after upgrading to version 1.6.3 on Windows 7 - WKB12311

Article number: 12311

SYMPTOMS

  • AAO Service Error
  • No communication to ACQUITY Classic system
  • Computer upgraded to Windows 7 64 bit
  • Analyst software upgraded to 1.6.3
  • Found: ACQUITY drivers v1.23b installed

ENVIRONMENT

  • AB Sciex Mass Spectrometer
  • Analyst 1.6.3
  • Windows 7 64 bit
  • ACQUITY Classic UPLC system

CAUSE

Incorrect ACQUITY UPLC drivers installed for the version of Analyst and Windows operating system.

FIX or WORKAROUND

  1. Remove ACQUITY UPLC drivers version 1.23b.
  2. Install appropriate drives for ACQUITY Classic UPLC system from the Waters Driver Pack 4 media or download.
  3. Install Waters Driver Pack 4 Analyst Update.

ADDITIONAL INFORMATION

 