"Padding is invalid and cannot be moved" error in Analyst software - WKB1012
Article number: 1012
SYMPTOMS
- ACQUITY UPLC Logon prompt appears after initializing Analyst software
- Entered Logon credentials
- "Padding is invalid and cannot be moved" error appears
- Cannot initialize inlet method due to error
ENVIRONMENT
- Analyst Software version 1.6
- ACQUITY UPLC System June 2011 Driver Pack (v1.50)
CAUSE
Corruption in a registry key associated with Analyst control of UPLC systems
FIX or WORKAROUND
- Start > Run, enter REGEDIT.
- Delete the registry key "SerializedObjects".
Windows XP:
Delete: HKeyLocalMachine\Software\Waters\Instruments\ACQ-AAO\ServiceSettings\SerializedObjects
Windows 7:
Delete: HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Waters\Instruments\ACQ-AAO\ServiceSettings\SerializedObjects
ADDITIONAL INFORMATION
If RUN is not found under the START menu, press the Windows Flag and R keys on the keyboard at the same time.
The deleted registry key will be recreated at the next logon to Analyst.