Error: System Code (71) - No more connections can be made to this remote computer at this time - WKB93563
Article number: 93563
SYMPTOMS
- The following error is reported in the NuGenesis Archive Agent activity log:
-
"System Code (71) - No more connections can be made to this remote computer at this time because there are already as many connections as the computer can accept."
-
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower
CAUSE
The folder being scanned has an "Advanced share" setting of 1 for "Limit the number of simultaneous users to".
FIX or WORKAROUND
- Open Folder Properties for the folder being shared to NuGenesis.
- Click on the Sharing tab.
- Click on the Advanced sharing button.
- Modify the "Limit the number of simultaneous user to:" setting back to the default of 20.
ADDITIONAL INFORMATION
In Empower, check the Empower Projects share settings.