LWRs are not transferred from NuGenesis Stability to LMS - WKB77878
Article number: 77878
SYMPTOMS
- One or more LWRs are pulled in the NuGenesis Stability module but are not transferred into LMS as new submissions
- The following message appears in the LMS manager.log file:
- No database entry for Formulation = Formulation Name
- The relevant Product entry in LMS does have a formulation with the same name given in the log file
ENVIRONMENT
- NuGenesis 9 LMS
- NuGenesis 8 LMS
- NuGenesis Stability
CAUSE
The formulation name in LMS includes an extra space.
FIX or WORKAROUND
- Select the formulation in LMS and click Copy into new formulation.
- Type the correct formulation name without extra blank spaces.
- Click OK.
- When the background job for LWR transfers next runs, it will find a match for the formulation and, providedthere are no other errors, it will create submissions in LMS for the LWRs.
ADDITIONAL INFORMATION
LMS allows blank spaces to be entered before and after formulation names. These extra spaces are removed when the formulation is transferred over to the Stability module, and when an LWR is pulled, the formulations will not match, resulting in this error.