Error: NuGenesis NG14419 Managed Store Error: Error while opening file - WKB67093
Article number: 67093
SYMPTOMS
- File in NuGenesis TE_Temp folder is not transferring to Managed Store location
- Error: NuGenesis NG14419 Managed Store Error
ENVIRONMENT
- NuGenesis 8 SR2
- NuGenesis 9
CAUSE
Managed Store location is a NAS with McAfee Virus scanning.
McAfee was scanning the file as NuGenesis was attempting to read the file.
FIX or WORKAROUND
- Create an exception rule for NuGenesis Managed Store location.
- The exception rule is required for .1 file extension.
ADDITIONAL INFORMATION
Raw data start with NG<ID>
Data Base CLOB start with NGC<ID>
NGD files start with NGD<ID>
id67093,