Form audit trail is not retained if form is recovered after an ELN.exe crash - WKB6585
Article number: 6585
SYMPTOMS
- A SmartBuilder form's audit trail report does not contain entries for data that was added to the form.
- The form was recovered by LMS following an LMS or Excel program crash.
ENVIRONMENT
- NuGenesis 9 LMS
- NuGenesis 8 LMS
- NuGenesis 8 ELN
CAUSE
The auto-recover file does not contain the form's audit trail information. As a consequence, any forms recovered from the Recover file do not have audit entries for all data changes made prior to the program crash.
FIX or WORKAROUND
- Defect reports INFLMS-9194, CRI-329, and INFLMS-2507 were filed for this issue for NG8 SR1, NG8 SR2, and NG9, respectively.
- Hofixes are available for the following versions:
- NuGenesis 8 ELN SR1
- NuGenesis 8 LMS SR2
ADDITIONAL INFORMATION