Error "Can't reach this page" appears in Internet Explorer when browsing to SDMS WebVision - WKB64423

Article number: 64423

SYMPTOMS

  • The following error message appears when browsing to SDMS WebVision:
    • Can't reach this page
      • Make sure the web address is correct
      • Search for this site on Bing
      • Refresh the page
  • The message "Page cannot be displayed" appears in Internet Explorer when loading the WebVision site
  • The SDMS Web Server's Hostname, used in the WebVision URL, cannot be contacted by name using Ping or any other method from the local PC  
  • The IP address of the SDMS Web Server responds using Ping
  • People can connect to Vision from other PCs using the same WebVision URL

ENVIRONMENT

  • NuGenesis 9 SDMS
  • NuGenesis 8 SDMS

CAUSE

The DNS server doesn't have an entry for the SDMS Web server.

FIX or WORKAROUND

  • Contact the IT department to have the DNS server's records updated so that it can resolve the SDMS Web server's name to an IP address.
  • A short-term fix is to modify the hosts file on the client machine so that it has an entry for the SDMS Web server. This solution is not recommended for long-term use because the Web server's IP address may change due to DHCP lease expiration.
    1. Open Notepad with Admin rights.
    2. Click File > Open and browse to: C:\Windows\system32\drivers\etc\hosts.
    3. Add a new line to the end of the file. Type the server's IP address, a space, and the server's hostname. Save and close the file.

ADDITIONAL INFORMATION

 