NuGenesis 8 LMS users cannot preview or restore in SDMS Project Browser - WKB6358
Article number: 6358
SYMPTOMS
-
Non-administrator users cannot preview or restore data using the SDMS Project Browser in NuGenesis LMS
ENVIRONMENT
- NuGenesis 8 SR2
- NuGenesis 8 FR1
- NuGenesis 8 SR1
CAUSE
Users have no privileges in SDMS to preview or restore data.
FIX or WORKAROUND
-
Log in to SDMS Administrator.
-
Click the project under Privileges.
-
Select the project to grant necessary privileges.
-
Click the VISION tab and select Preview Report and Restore Data.
-
Save changes.
ADDITIONAL INFORMATION